IKT Measurement Data / VME-Analysis
Location/Site:
IKT Laboratory, Winterthur/CH
Date:
Comments/Conclusions
The critical perturbations at 15h10 (red) represent an
unacceptable environment for the micro balance.
The critical permanent perturbations over a 15 minutes
time period show amplitudes over 2000µm/s.
As expected by experience, values over about 2000µm/s
get critical.
Such perturbations have to be analyzed and to be
reduced or removed by
damping, geotechnical or other
organisational measures.
Red arrows: low frequency strong permanent shaking perturbation (on the
granite table).
Green arrows: single delta peaks
perturbation (hit on the granite table in the z-direction)
Last update: 2010-03-19